3502 EZ Slide Surgical Table

Starting date:
September 28, 2016
Posting date:
March 31, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62832

Reason

The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the table's elevation column slide cap to become detached from the elevation column. This could result in the loss of control of the table top and injury to the patient or staff.

Affected products

3502 EZ Slide Surgical Table

Lot or serial number

  • 3502-SUT-20144-0155
  • 3502-SUT-20147-0222
  • 3502-SUT-20148-0234
  • 3502-SUT-2014Z-0313
  • 3502-SUT-20151-0342
  • 3502-SUT-20154-0361
  • 3502-SUT-20154-0362
  • 3502-SUT-20156-0392
  • 3502-SUT-20156-0393

Model or catalog number

  • 3502

Companies

Manufacturer

Skytron LLC