This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

348 System - Gas Cartridge Pack

Starting date:
May 25, 2017
Posting date:
June 13, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63610

Reason

Siemens Healthcare has received a number of reports of gas cartridges failing to install on the RapidLab 348 and RapidLab 348EX instruments due to defective threads on the gas cartridges. The issue is intermittent and not all gas cartridges are affected.

Affected products

348 System - Gas Cartridge Pack

Lot or serial number

5260xx through 7020xx

Model or catalog number

  • 10309768
  • 105070

Companies

Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

UNITED STATES