2.5mm Lifeport Adapters and Patient Circuit Kits
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
2.5mm Lifeport Adapters 10-Pack |
22042320 |
9025 |
Patient Circuit Kits |
22052381 |
938 |
Issue
Bunnell has received reports of 2.5 mm Lifeport adapters with deformations at the patient end tip. These deformations may result in a narrowing of the effective internal diameter of the Lifeport. An evaluation of returned adapters has confirmed the presence of deformations. Bunnell incorporated is voluntarily recalling a limited number of 2.5mm Lifeport adapters (lot number 22042320) for use with the lifepulse high frequency ventilator. The affected 2.5mm Lifeport adapters were packaged in these products: catalog #9025 (2.5 mm Lifeport adapters 10-pack) and catalog #938 (patient circuit kits).
Recall start date: November 2, 2022
Additional information
Details
Bunnell Inc
436 Lawndale Drive, Salt Lake City, Utah, United States, 84115
Get notified
Receive emails about new and updated recall and safety alerts.