This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
2008T Hemodialysis Machine (2014-05-28)
- Starting date:
- May 28, 2014
- Posting date:
- July 3, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40289
Recalled Products
2008T Hemodialysis Machine
Reason
Several of the affected actuator test boards were found to not reliably detect a disconnected ultrafiltration (UF) pump.
Affected products
2008T Hemodialysis Machine
Lot or serial number
Not applicable
Model or catalog number
190234
Companies
- Manufacturer
-
Fresenius USA Inc.
4040 Nelson Avenue
Concord
94520
California
UNITED STATES