Health product recall

1688 4K PENDULUM CAMERA HEAD WITH INTEGRATED COUPLER

Last updated

Summary

Product
1688 4K PENDULUM CAMERA HEAD WITH INTEGRATED COUPLER
Issue
Medical devices - Software defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or Serial Number

Model or catalog number

1688 4K PENDULUM CAMERA HEAD WITH INTEGRATED COUPLER

More than 10 numbers, contact manufacturer.

1688310130

Issue

Through investigation following a product compliant (November 11, 2021), Stryker has identified a software bug in the 1688 camera control unit (CCU) that can cause the image on the monitor to flip upside down (only) when used in conjunction with the 1688-310-130 Pendulum Camera Head.

Recall start date: Jan 28, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

STRYKER CANADA ULC

2 MEDICORUM PLACE, WATERDOWN, ON

CANADA, L8B 1W2

 

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63908

Get notified

Receive notifications for new and updated recalls and alerts by category.