This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

1,25-Dihydroxy Vitamin D Ria

Starting date:
March 14, 2017
Posting date:
April 13, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63014

Reason

In a small number, approximately 6%, of 1,25-dihydroxy vitamin d ria kits containing calibrators D lot 34649 and E lot 34650, similar CPM values for both calibrators are obtained. These abnormal results lead to an incorrect calibration curve fit.

Affected products

1,25-Dihydroxy Vitamin D Ria

Lot or serial number

  • J40007
  • J40258
  • J40662
  • J40785
  • J40856

Model or catalog number

  • AA-54F1
  • AA-54F2

Companies

Manufacturer

Immunodiagnostic Systems Limited

10 Didcot Way, Boldon Business Park,

Boldon, Tyne and Wear,

NE35 9PD

UNITED KINGDOM