Health product recall

12-Instrument Sterilization Tray (2014-04-25)

Starting date:
April 25, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39479

Recalled products

  1. 12-Instrument Sterilization Tray 

Reason

Gravity sterilization parameters provided in the 12 instrument sterilization reprocessing guide (1000400920 Revision D) did not fully consider all worst-case scenarios during validation testing.

Affected products

A. 12-Instrument Sterilization Tray 

Lot or serial number

All lots  

Model or catalog number
  • 242-000-012
Companies
Manufacturer

Stryker Endoscopy

5900 Optical Court

San Jose

95138

California

UNITED STATES