10Fore™ Hemostasis Valve

Last updated

Summary

Product
10Fore™ Hemostasis Valve
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

10Fore™ Hemostasis Valve

H3098829

MAPTEN452

10Fore™ Hemostasis Valve

H3111162

MAPTEN452

10Fore™ Hemostasis Valve

H3098831

MAPTEN4

10Fore™ Hemostasis Valve

H3111155

MAPTEN4

10Fore™ Hemostasis Valve

H3152503

MAPTEN4

10Fore™ Hemostasis Valve

H3155611

MAPTEN4

10Fore™ Hemostasis Valve

H3186901

MAPTEN4

10Fore™ Hemostasis Valve

H3076189

MAPTEN42

10Fore™ Hemostasis Valve

H3111156

MAPTEN42

10Fore™ Hemostasis Valve

H3155614

MAPTEN42

10Fore™ Hemostasis Valve

H3173674

MAPTEN42

10Fore™ Hemostasis Valve

H3152504

MAPTEN42

Issue

Merit Medical became aware of a manufacturing quality concern during routine QA inspection of a 10Fore™ Hemostasis Valve work order in February 2025. Initial defect assessments determined the observed damage to the silicone quad ring was due to the manual nature of operator handling and subsequent placement/positioning of the quad ring into the corresponding y body. If the ring is not gently manipulated and/or not fully seated in its intended position during the pressing process it can become deformed or otherwise damaged.

Recall start date: July 25, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Merit Medical Systems Inc.

1600 West Merit Parkway, South Jordan, Utah, United States, 84095

Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-77789

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe