0.9% Sodium Chloride Injection, USP and Lactated Ringer’s Injection, USP - Potential Leak During Pressure Infusion
Summary
See Key Messages below
Affected products
Product Name |
Package Size |
Product Code |
Market Authorization Holder |
DIN |
Lot Numbers/ Expiration Dates |
---|---|---|---|---|---|
0.9% Sodium Chloride Injection, USP |
500 mL |
JB1323 |
Baxter Corporation |
00060208 |
See Appendix A |
Lactated Ringer’s Injection, USP |
500 mL |
JB2323 |
Baxter Corporation |
00061085 |
See Appendix A |
Issue
Intravenous bags of Baxter’s 0.9% Sodium Chloride Injection, USP 500 mL and Lactated Ringer’s Injection, USP 500 mL from certain lots have the potential to leak when administered under pressure infusion due to a weak seal formation.
Affected lots are not being recalled to prevent a shortage of these medically necessary products.
Leakage from the bags could:
- lead to contamination of the solution and may result in a serious blood infection. Patients with an infection of the blood may develop a fever, low blood pressure, or an altered level of consciousness. Blood infections may be life-threatening in both adults and children.
- result in an inaccurate amount of medication being delivered when other medications have been mixed into the 0.9% Sodium Chloride Injection, USP 500 mL and Lactated Ringer’s Injection, USP 500 mL as part of routine treatments.
- expose healthcare professionals or patients handling the bags to hazardous medications (e.g., cytotoxic chemotherapeutic drugs) when these medications have been added to the bags.
Audience
Healthcare professionals including physicians, nurses, pharmacists, and other medical and support personnel including home healthcare agencies involved in the administration and handling of 0.9% Sodium Chloride Injection, USP 500 mL and Lactated Ringer’s Injection, USP 500 mL products.
Key messages
- Intravenous bags of Baxter’s 0.9% Sodium Chloride Injection, USP 500 mL and Lactated Ringer’s Injection, USP 500 mL from certain lots have the potential to leak when administered under pressure infusion.
- Leakage from the bags could:
- lead to contamination of the solution which may result in a serious blood infection.
- result in an inaccurate amount of medication being delivered when other medications have been mixed into the 0.9% Sodium Chloride Injection, USP 500 mL and Lactated Ringer’s Injection, USP 500 mL.
- expose healthcare professionals or patients handling the bags to hazardous medications (e.g., cytotoxic chemotherapeutic drugs) when these medications have been added to the bags.
- Healthcare professionals are advised of the following:
- Do NOT administer products from affected lots (see Appendix A) under pressure infusion.
- Regardless of the lot, prior to administration, tear the overwrap (outer bag) down the side at the slit and remove the inner bag. Inspect the bag for low fill volume and for leaked fluid in the space between the overwrap and inner bag. Check for minute leaks by squeezing the inner bag firmly.
- If any leakage is observed, the bag should be discarded due to the potential risk of contamination.
- Monitor for leakage from the bag during treatment and stop using if leakage is observed. In the affected lots, leaks are more likely to be found in the area highlighted in the photo in Appendix B.
- Gloves should be worn when handling the bags to avoid exposure to any hazardous medications that may have been added to the bags.
Background
0.9% Sodium Chloride Injection, USP (DIN 00060208) is indicated as a source of water and electrolytes. It can also be used as a vehicle or diluent for compatible products for parenteral administration, or for use as a priming solution in hemodialysis procedures.
Lactated Ringer’s Injection, USP (DIN 00061085) is indicated as a source of water and electrolytes or as an alkalinizing agent.
Baxter identified the cause of the leakage of the 500 mL intravenous bags as a weak seal formation due to inappropriate manufacturing equipment set-up. Baxter has implemented corrective actions to prevent reoccurrence. There have been no reports of adverse events to date for the affected lots.
Products from the affected lots were distributed by Baxter Corporation in Canada between May 21, 2021 and March 01, 2022.
Information for consumers (home patients)
Do NOT use the bags from affected lots in treatments where pressure is applied to the bags, as fluid may leak out.
If you see any fluid leakage from the bags before or during treatment, stop using the product as there may be a potential risk of contamination and consult your healthcare professional.
Gloves should be worn when handling the bag to avoid exposure to any toxic drugs (such as some cancer drugs) that may have been added to the fluid in the bags.
If you use these products with pressure, and you have product from an affected lot, do NOT use the bags and consult your healthcare professional to obtain a replacement product as soon as possible.
Consult with your healthcare professional if you have used product from an affected lot, or are unsure and have questions or concerns about your health.
Information for healthcare professionals
Bags of 0.9% Sodium Chloride Injection, USP 500 mL and Lactated Ringer’s Injection, USP 500 mL from the affected lots should NOT be used in pressurized intravenous lines due to the potential for product leaks.
Regardless of the lot, prior to administration, tear the overwrap (outer bag) down the side at the slit and remove the inner bag. Inspect the bag for low fill volume and for leaked fluid in the space between the overwrap and inner bag. Check for minute leaks by squeezing the inner bag firmly
If any leakage is observed, the bag should be discarded due to the potential risk of contamination.
Monitor for leakages from the bag during treatment and stop using if leakage is observed. Potential leakages may not be easily identified without the application of pressure over time. In the affected lots, leaks are more likely to be found in the area shown in Appendix B.
Gloves should be worn when handling the bags to avoid exposure to any hazardous medications that may have been added to the bags.
Action taken by Health Canada
Health Canada is communicating this important safety information to healthcare professionals and Canadians via the Recalls and Safety Alerts Database on the Healthy Canadians Web Site. This communication will be further distributed through the MedEffect™ e-Notice email notification system.
Health Canada has directed Baxter Corporation, the market authorization holder for 0.9% Sodium Chloride Injection, USP 500 mL and Lactated Ringer’s Injection, USP 500 mL, to ensure that they inform all impacted healthcare professionals, patients and personnel administering or handling these products.
Health Canada is monitoring the company’s implementation of any additional corrective and preventative actions. If additional safety information is identified, Health Canada will take appropriate action and inform Canadians as needed.
Report health or safety concerns
Health Canada’s ability to monitor the safety of marketed health products depends on healthcare professionals and consumers reporting adverse reactions and medical device incidents. Any case of product-related adverse reactions, including blood infections, or other serious or unexpected adverse reactions in patients receiving 0.9% Sodium Chloride Injection, USP 500 mL and Lactated Ringer’s Injection, USP 500 mL should be reported to Baxter Corporation or to Health Canada.
Baxter Corporation
7125 Mississauga Rd
Mississauga, ON L5N 0C2
To correct your mailing address or fax number, contact Baxter Corporation at fca_canada@baxter.com
You can report any suspected adverse reactions associated with the use of health products to Health Canada by:
- Calling toll-free at 1-866-234-2345; or
- Visiting MedEffect Canada's Web page on Adverse Reaction Reporting for information on how to report online, by mail or by fax.
For other health product inquiries related to this communication, contact Health Canada at:
Health Product Compliance Directorate, Regulatory Operations and Enforcement Branch
E-mail: hpce-cpsal@hc-sc.gc.ca
Telephone: 1-800-267-9675
Original signed by
Azmina Kanji
Manager, Quality Systems
Baxter Corporation - Canada
Appendix A: Affected Lots
Product Name: 0.9% Sodium Chloride Injection, USP
Package Size: 500 mL
Product Code: JB1323
Market Authorization Holder: Baxter Corporation
DIN: 00060208
|
Lot Number |
Expiration Date |
1 |
W1I14B0F |
DEC22 |
2 |
W1I15B0 |
DEC22 |
3 |
W1I23B0 |
DEC22 |
4 |
W1I24B0 |
DEC22 |
5 |
W1I30B0 |
DEC22 |
6 |
W1J07B0 |
JAN23 |
7 |
W1J20B0 |
JAN23 |
8 |
W1J22B0 |
JAN23 |
9 |
W1J25B1 |
JAN23 |
10 |
W1J26B0 |
JAN23 |
11 |
W1J27B1 |
JAN23 |
12 |
W1J29B0 |
JAN23 |
13 |
W1K02B0 |
FEB23 |
14 |
W1K03B0 |
FEB23 |
15 |
W1K19B0 |
FEB23 |
16 |
W1K25B0 |
FEB23 |
17 |
W1K25B0F |
FEB23 |
18 |
W1L02B0 |
MAR23 |
19 |
W1L03B0 |
MAR23 |
20 |
W1L09B0 |
MAR23 |
21 |
W1L10B0 |
MAR23 |
22 |
W1L10B0B |
MAR23 |
23 |
W1L11B0 |
MAR23 |
24 |
W1L16B0 |
MAR23 |
25 |
W1L17B0 |
MAR23 |
26 |
W2A13B0 |
APR23 |
27 |
W2A14B0 |
APR23 |
28 |
W2A19B1 |
APR23 |
29 |
W2A20B0 |
APR23 |
30 |
W2A21B0 |
APR23 |
31 |
W2A22B0 |
APR23 |
32 |
W2A26B0 |
APR23 |
33 |
W2A29B0 |
APR23 |
34 |
W1E05B0 |
AUG22 |
35 |
W1E06B0 |
AUG22 |
36 |
W1E11B0 |
AUG22 |
37 |
W1E18B0 |
AUG22 |
38 |
W1E21B0 |
AUG22 |
39 |
W1E27B0 |
AUG22 |
40 |
W1F03B0 |
SEP22 |
41 |
W1F04B0 |
SEP22 |
42 |
W1F11B0 |
SEP22 |
43 |
W1F12B0 |
SEP22 |
44 |
W1F18B0 |
SEP22 |
45 |
W1F22B0 |
SEP22 |
46 |
W1F23B0 |
SEP22 |
47 |
W1F25B0 |
SEP22 |
48 |
W1F28B0 |
SEP22 |
49 |
W1F29B0 |
SEP22 |
50 |
W1F30B0 |
SEP22 |
51 |
W1G14B0 |
OCT22 |
52 |
W1G15B0 |
OCT22 |
53 |
W1G16B0 |
OCT22 |
54 |
W1G16B0B |
OCT22 |
55 |
W1G20B0 |
OCT22 |
56 |
W1G21B0 |
OCT22 |
57 |
W1G29B0 |
OCT22 |
58 |
W1G30B0 |
OCT22 |
59 |
W1H13B0 |
NOV22 |
60 |
W1H19B0 |
NOV22 |
61 |
W1H20B0 |
NOV22 |
62 |
W1H26B0 |
NOV22 |
63 |
W1H27B0 |
NOV22 |
64 |
W1I01B0 |
DEC22 |
65 |
W1I03B0 |
DEC22 |
66 |
W1I09B0 |
DEC22 |
67 |
W1I10B0 |
DEC22 |
68 |
W1I14B0 |
DEC22 |
Product Name: Lactated Ringer’s Injection, USP
Package Size: 500 mL
Product Code: JB2323
Market Authorization Holder: Baxter Corporation
DIN: 00061085
|
Lot Number |
Expiration Date |
1 |
W1H04B0 |
NOV22 |
2 |
W1I17B0 |
DEC22 |
3 |
W1J28B0 |
JAN23 |
4 |
W1K26B0 |
FEB23 |
5 |
W1L22B0 |
MAR23 |
6 |
W2A27B0 |
APR23 |
7 |
W2A27B0S |
APR23 |
8 |
W1E14B0 |
AUG22 |
9 |
W1F24B0 |
SEP22 |
Appendix B: Area or Location of Potential Leakage
In the affected lots, leaks are more likely to be found in the area shown below:
If any leakage is observed from the bags before or during treatment, stop using the product.