0180/OMNERA 400T (2021-05-17)
- Starting date:
- May 17, 2021
- Posting date:
- June 17, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75811
Last updated:
2021-06-17
Affected Products
0180/OMNERA 400T
Reason
Cleaning of the maneuver handle or the maneuver display with excessive amount of disinfectants containing certain components pose a risk of causing a short circuit due to ingress of liquid, which in turn could cause uncontrolled up- or down movement of the overhead tube crane (OTC).
Affected products
0180/OMNERA 400T
Lot or serial number
All lots.
Model or catalog number
180
Companies
- Manufacturer
-
ARCOMA AB
ANNAVAGEN 1
VAXJO
SWEDEN